PATIENT MONITOR - SAADAT - MALAYSIA
PATIENT MONITOR
MERK : SAADAT - MALAYSIA
MODEL : ALBORZ B9
Specifications :
DISPLAY
Displaying Color LED 1366 x 768,18.5"
Printing Speed 12.5, 25, 50 mm/sec
Operation Method Rotary Knob & Membrane
ECG :
Leads : 3/5 leads
Dynamic Range : ± 5 mV
Leakage Current : < 10µA
Lead Off Cureent : < 90 nA
Gain : 4, 2, 1, 1/2, 1/4, Auto
Calibration : 1mV, 0.5 sec
Filters : "MONITOR" ( 0.5 - 24Hz )
"NORMAL" ( 0.5 - 40 Hz )
"EXTENDED" ( 0.05 - 100 Hz )
CMRR : > 98dB
Internal Noise : < 30 µV RTI
Input Impedance : > 5 Mohm
QRS Detection : Duration : 40 to 120 msec
Amplitude : 0.25 to 5 mV for Adult
0.2 to 5 mV for Neonate
Heart Rate Range : 15 - 300 bpm for Adult
15 - 350 bpm for Neonate
Accuracy : ±1% or 2 bpm
Tall T-wave : Reject up to 1.2 mV Amp.
Pace Detection / Rejection : Duration : 0.1 - 2 msec
Amp : ± 2 to ± 700 mV ( Without over/undershoot )
Reject From Heart Rate Counter
Re-insert into ECG to display on screen
Protection : Defibrillator and Electro surgery
Standards : ANSI / AAMI EC - 13
ARRHYTHMIA ANALYSIS :
Type : ASYS, VFIB, VTAC, RUN, AIVR, COUPLET, BIGEMINY, TRIGEMINY,
TACHY, BRADY, PAUS, FREQUENT PVCs
Learning : Rapid Learning : only 20 Second Required or Recognition of
Dominant Rhythm
Methode : Real Time Arrhythmia Detction with Innovative Feature.
Memory : Capability of storing the latest 80 ARR event.
( waveform and parameters )
ST ANALYSIS :
Display resolution : 0.01 mV
D E S C R I P T I O N
Page 2
NO D E S C R I P T I O N
Measurement Range : -2 mV to +2mV
Alarm Range : -2 mV to +2 mV
Features : User Adjustable Isoelectric and ST Point Trending of ST Values
Update period : 5 sec.
Masimo SET SPO2
Method : 2 wave Length Pulse Wave Type
Range : 50% to 100%
Accuracy : ± 2% (SPO2 71 ~ 100%)
: ± 3% (SPO2 50 ~ 70%)
Pulse Rate Range : 25-250 bpm
NIBP :
Measurement methode : Oscillometric
Measurement mode : Manual/ Automatic
( intervals between 5min - 24hour ) / STAT
Measurement time : 18-26 sec ( excluding cuff pressurization time )
Measurement Range : Adult : SYS 25 ~ 250 mmHg
DIA 10 ~ 220 mmHg
MAP 15 ~ 250 mmHg
Neonate : SYS 25 ~ 135 mmHg
DIA 10 ~ 110 mmHg
MAP 15 ~ 125 mmHg
Pressure Transducer accuracy : ±3 mmHg full range
Initial Inflation Target : Adult 150 mmHg , Neonate 70 mmHg
Overall System Efficacy : Meet ANSI / AAMI SP-10/1992
Memory : Up to 100 Records
RESPIRATION :
Method : Impedance
Base Resistance : 250 - 1250 Ohm
Dynamic Range : 0.2 - 2 Ohm
Breath Rate Range : 6 - 150 BrPM
TEMPERATURE :
Probe Type : YSI 700 Compatible
Range : 0 - 50 °C
Accuracy : ±0.2 °C
ALARM :
Sources : Error messages,all other Parameter Limits
Alarm On / Off : Selectable for All Parameters
Alert : Blinking on Display,
Volume Selectable Audio Alarms,Light indicator
INPUT / OUTPUT
Network : Ethernet LAN
Page 3
NO D E S C R I P T I O N
Connection : Up to 16 BEDs to one CENTRAL
GENERAL :
Safety : Based on IEC 60601-1
Protection : Against Electro surgery and Defibrillator
AC power : 90 - 260 VAC,50/60 Hz
Internal Battery : Li-ion or (Sealed Lead Acid)
Rechargeable 12 V, 5.0 AH (3.3AH)
Charge time 16 Hours
Usage more than 4 hours ( 2 hours )
DC Power Plug : 12 - 14V - 3A / ( 6A with recorder )
Dimension ( Cm ) : 45 (W) x 35 (H) x 16 (D)
Nett weight : 5.5 to 7.5 Kg
RECORDER
Model : Thermal Printer
Printing Speed: 12.5, 25, 50 mm/sec.
Paper : 58 mm by 100 foot roll
ENVIRONMENTAL :
Temperature : Operating 5 to 40°C
Storage : -20 to 60°C
Humidity : 20 - 90% ( Non condensing )
Altitude : -200 to 3000m
All specifications are subject to modifications and improvements without prior notice